Digital Weighing Indicator
SI 4100
G
TGX^GTG
5-3. Simulation Calibration Mode (Calibrate without Test weight)
Through this
“Simulation Calibration Mode”
you can make simple calibration without Test weight
.
This calibration mode uses “Load cells’ max. capacity” and “Max. Output Rate(mV)”, the weight adjustment
degree might be less than “Test weight Calibration”.
The guaranteed resolution of this “Simulation Calibration” is 1/3,000.
Remove
“CAL-BOLT”
on the Front panel and press
“CAL - LOCK S/W”
inside.
Then, you can enter the Calibration Mode with SET-CAL and press
key to enter “
Simulation
Calibration Mode”
with "CELL CAL" and start calibration mode with pressing
key.
1. At normal mode, remove “CAL-BOLT” on the Front panel
1123
2. And press “CAL - LOCK S/W” inside.
Check the “SET-CAL. Massage on display.
5et-Cal
3. Please press
key, To start Simulation Calibration Mode
CellCal
G
ଖ
For the save the each step, press
key, for the cancel or move back, press
key.
G
4. Press
key to enter calibration mode.
After displaying “CAPA”, you may set Max capacity..
C999999
5. After input Max capacity of your weighing scale (at the label),
press
key
Ex) Load cell CAPA : 30kg, division : 0.01
Input 3000
C 3000