- 34 -
Digital Filter setting
F06 23 AB
A : Frequency Filter setting value (0~3)
(0 : about 200Hz/sec, 3 : about 500Hz/sec)
B : Buffer Filter setting value (1~9)
If “B” set value is fixed,
“A” set value is large, the
indicator will response
more sensitive.
※
This function is related with “Zero Tracking Compensation Range setting(F04)”
※
If there is strong vibration near weighing part, please set “A” with large value.(If the “B” set value is
fixed.)
Zero /Tare key Operation mode selection
●
0
Activate
when
“Steady” condition, only
F07
1
Always
activated
Zero key Operation Range selection
0
Activated within 2% of Max. Capacity
1
Activated within 5% of Max. Capacity
●
2
Activated within 10% of Max. Capacity
F08
3
Activated within 20% of Max. Capacity
Tare key Operation Range selection
0
Activated within 10% of Max. Capacity
1
Activated within 20% of Max. Capacity
●
2
Activated within 50% of Max. Capacity
F09
3
Activated within 100% of Max. Capacity
“Key TARE” Operation Selection
●
0
Key TARE Function Not Use.
F10
1
Key TARE Function Use
※
Key TARE Function : You can set TARE value with No. keys, without weight.
※
To input Key TARE value(F10-01 setting),
Press
“TARE”
key, and input
“target TARE value”
with No. keys and press
“ENTER”
.