11 Faults and problems
│
57
Information
Explanation/corrective action
List empty
The list of protocols/errors does not
contain any entry because no de-
vice inspection or error has yet been
saved.
Memory full. Old
protocols will be
overwritten.
All memory slots in the protocol
memory are in use. The oldest pro-
tocols are overwritten by the new
protocols (ring memory).
– Create space in the protocol
memory using
GasCom
: Read
protocols >
Download
>
Delete
New test gas. Currently
assigned: …
New test gas connected.
Check whether the new test gas cor-
responds to the currently assigned
test gas.
–
If necessary, reassign test gas
(
Settings
>
Gas assignment
).
No device (PM)
detected.
Message appears in
Manual
operat-
ing mode when a device inspection
should be started even though no
device is inserted.
– Insert the device.
Test gas pressure too
low.
Test gas can or test gas bottle is
almost empty.
– Connect new test gas.
Test gas can insufficiently screwed
on.
– Screw on the test gas can tighter.
“Timer” operating mode
not possible because no
days are specified.
– In
Settings
>
Timer
, set the
Day
of the week
and the
Start time
.