7 Device inspections
│
41
7.7
Performing the device inspection
7.7.1
Manual operating mode
Fig. 30:
Manual
operating
mode,
Device inspection menu
The menu item
All
is displayed
at the top position because more
than one device inspection is
due.
The test set is switched on.
Manual
operating mode is se-
lected.
1. Check the device housing for external damage.
3
2. Place the device in the test set.
3. Select
All
or an application.
4. Press the Enter key. The device inspection starts.
The automated part of the device inspection is complete when
the following message appears:
Device condition: Display
inverted?
5. Check the device condition.
A series of questions appears. Answer them.
a)
Device condition: Display inverted?
Does the device display show all pixels in reverse colour
assignment
4
(fig. 31)?
3
The inspection is required before insertion so that the device does not
have to be removed from the test set for the
Device condition
subtest. If
you remove the device during a running device inspection, the inspection
will be cancelled.
4
White pixels are displayed black, black pixels are displayed white.