38
│ 7 Device inspections
Variants of the device inspections for the warning
application (standard and ECO)
In
Manual
operating mode, the device inspection for the
Warn-
ing application
can be performed in two variants. The variants
differ in whether the indication accuracy is tested and whether a
bump test is performed.
Device inspection
(
Manual
operating mode /
Warning
application)
Standard device inspection
Subtests:
−
Pump
−
Zero point
−
Indication accuracy
−
Device condition
Name of the menu item:
Warning
ECO
device inspection
Subtests:
−
Pump
−
Zero point
−
Bump test
−
Device condition
Name of the menu item:
Warning ECO
Fig. 27:
Manual
operating mode,
Warning
application – variants of de-
vice inspection with associated subtests
7.3.3
Automatic and Timer operating modes: Special features
●
In the
Automatic
and
Timer
operating modes, the test set
performs all the device inspections that are due automatically
one after the other.
● However, the
Device condition
subtest requires user interac-
tion with the test set.