SAMPLE SYSTEM
INTEGRATION
The SERVOTOUGH Oxy
is a precision analy�cal device, suitable for a wide variety of
indu
strial applica�ons. Though every effort has been taken to ensure that the
analyser is highly robust to differing sample stream composi�ons, it should be noted
that if the requirements of Sec�on 2.4
and
Sec�on 2.5 are not met then the analyser
can be damaged and the measurement invalidated.
9.1
Sample
Systems
The analyser will be damaged if condensa�on is allowed to form within the
measurement cell or pipe work. It is important that the sample gas dew points
specified in Sec�on 2.4 are maintained at all �mes.
Note:
It is recommended that sample systems contain a flow monitoring device on the outlet
of the analyser, ideally with a status indication communicating back to the control
room, or that the analyser is fitted with a Servomex Flowcube flow sensor option. Any
failure in the sample flow can then be quickly diagnosed and corrective action taken.
Note:
In the event of power failure to the analyser or the sample system, condensation can
form within the sample pipework if the sample gas has a dew point below that of the
ambient temperature.
It is recommended that if there is a power failure that the sample system contains
sufficient protection and interlocks on coolers and "hot" sample elements to prevent
gas flow though the analyser until these are at working temperature.
If the sample gas is considered corrosive, then it is recommended that during a power
failure or while the sample system is not at operational conditions, that instrument air
is used to purge the measurement pipework of the analyser.
9.2
Corrosive
Purge
Failure of the corrosive purge could result in damage to the electronics within the
sample/control compartment, which could result in an analyser failure.
The corrosive purge gas must comply with the requirements of Sec�on 2.5. If the
purge gas is not dry or contains other contaminates, i.e. oil vapours, then this can
result in damage to the electronics within the sample/control compartment, which
could result in an analyser failure.
Summary of Contents for SERVOTOUGH Oxy
Page 2: ...This page inten onally blank...
Page 4: ...This page inten onally blank...
Page 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Page 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...