7.
GENERAL
OPERATION
Sample and calibra�on gases must be as specified in Sec�ons 2.4
. If the
pressure/flow rates are outside the ranges specified in Sec�ons 2.4
, you must
regulate the gases externally, before they enter the analyser
.
7.1
Checking the relay signal
outputs
Note:
Gas selection relays
(Sample, Cal lo, Cal hi) are not effected during the signal relay
override as this could
cause hazardous gas flows.
If required, use the following procedure at any �me to perform a check on the outputs of the signal
relays fited to the analyser
:
1.
Selec
t:
Service
Relay
2.
shows the relay state (Energised or Deenergised) that the relay signals outputs
will be set to when the override is ac�ve. Edit if required.
Figure
88 -
The Relay override state screen
Note:
When a relay is energised, the N/C
-
Common contacts will be open, and the N/O
-
Common contact
s will be closed. The converse is true when the relay is deenergised.
3.
Scroll to the relay override
screen.
4.
Select
the "Yes" op�on, then press the
so� key. The relay outputs will now be set
to the selected override state, and you can use your control/monitoring equipment
(connected to the analyser) to monitor the relay signal outputs
Figure
89 -
The Relay override ac�on screen
Note:
The relay signal outputs freeze at the selected override signal state as long as the
‘Override action’ screen is displayed. As soon as another screen is displayed, the relay
signal outputs will be updated to reflect the corresponding alarm, fault and range
states.
Summary of Contents for SERVOTOUGH Oxy
Page 2: ...This page inten onally blank...
Page 4: ...This page inten onally blank...
Page 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Page 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...