![Servomex SERVOTOUGH Oxy Operator'S Manual Download Page 116](http://html1.mh-extra.com/html/servomex/servotough-oxy/servotough-oxy_operators-manual_1247065116.webp)
104
© Servomex Group Limited. 2020
01910001B_8
Register
Name
Comments
101 - 102
Measurement
As seen on the measurement display. Floa�ng
point number.
103 -
104
Filtered
measurement
Basic measurement, filtered. Floa�ng point
number.
105 - 106
Pressure
compensated
measurement
Filtered and pressure compensated
measurement. Floa�ng point number.
107 - 108
PMR
Primary measurement range
109
Warming state
0 = Not warming, 1 = Warming
110 - 112
Formula
Gas formula. ASCII text, 5 characters max,
terminated by a zero.
113 - 116
Units
Measurement units. ASCII text, 7 characters
max,
terminated by a zero.
117
Auto val state
0 = Idle, 1 =
Pre-
warning, 2 = Iner�ng, 3 =
Flushing, 4= Valida�ng/Calibra�ng
118
Auto val gas
0 = Low gas, 1 = High gas, 2 = Sample gas
119
Auto val finishing
0 = Not finishing, 1 = Finishing
120
Auto val fail state
0 = Auto valida�on OK, 1 = Auto valida�on
failed
121
Ac�ve analogue
output range
0 = Range 1, 1 = Range 2
122
Flow
Current flow through the transducer in percent
A.10
System fault mapping
Read-
only access to system fault informa�on will be provided in a block of 100 registers. These can be
accessed
as discrete inputs using func�on code 02 or as input registers using func�on code 04. Reading
them as discrete inputs provides a simple bit result for each, 0 for off and 1 for on. Reading them as
input registers provides informa�on on their NAMUR status, as follows:
0 =
off,
1 =
fault,
2 = maintenance
required,
3 = service in
progress.
The fault, maintenance required, and service in progress faults provide a summary of the other states
and can only be off or on. The registers are assigned as follows:
Summary of Contents for SERVOTOUGH Oxy
Page 2: ...This page inten onally blank...
Page 4: ...This page inten onally blank...
Page 20: ...8 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 28: ...16 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 30: ...18 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 41: ...01910001B_8 Servomex Group Limited 2020 29 Figure 12 Fixing dimensions mm...
Page 82: ...70 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 98: ...86 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 110: ...98 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 112: ...100 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 126: ...114 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 130: ...118 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 132: ...120 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 140: ...128 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 142: ...130 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...
Page 144: ...132 Servomex Group Limited 2020 01910001B_8 This page inten onally blank...