SIE-CNTEP-1100002104-2014
LCFS
AV6418 OTDR
107
OTDR test port reliably.
Ø
Enter OTDR test condition setup interface. (See P27 for
the detail of the operation )
Ø
Use / to move active cursor ( ) to
“
Mode
”
option or press
“
Mode
”
option directly through touchscreen to select the
OTDR test mode:
Real-Time Test
or
Average Test
or
Dead
Zone Test.
Ø
Operate it the same as above, set test
Wavelength,
Pulse Width, Att. IOR, Helix
and all
Threshold
. When the
test mode is average test or dead zone test, the average
times should be set.
Ø
Confirm OTDR parameter setting and return to main
screen. Press key, the instrument will scan optical fiber to be
tested according to the set test mode and test parameter
and display the trace on the operating interface.
Ø
Press key to stop sweep test at any time. If the test
mode is average test or dead zone test, the average times
displayed in the trace displayer area is equal to the set
average times, the instrument will automatically stop testing.
Ø
Press
【
Analyse
】
menu item on the main menu bar, the
instrument will automatically analyze the trace and give
event table according to the set event threshold. If it is set
“
Auto analyse after sampling
”
, when stopping test, the
instrument will automatically analyze the trace.
Ø
It can also set the proper mark point position through
setting test item, check the distance and loss information of
the optical fiber to be tested on the trace and result display
area.
Página No.119 LCFS
CONVALIDACIÓN DE ERRORES DEL PROCESO:
“ADQUISICIÓN DE OTDR (OPTICAL TIME DOMAIN REFLECTOMETER) – REDES GPON/