If the fiber link you are measuring is very short, such like
less than 100m, it is suggested to adopt below test mode
and test condition:
Please set: average test mode, range less than 800m, pulse
width less than 80ns and attenuation of 15dB or 20dB, and
set certain average time.
Setting an attenuation of 15dB may obtain a test trace with
better signal to noise ratio, which is of advantage to make
trace analysis. Setting an attenuation of 20dB may obtain a
smaller test dead zone, but the signal to noise ratio will be
relatively poor, there requires setting a larger average time
to obtain a trace of better noise ratio.
Ø
Near end of the test trace has severe tailing.
Try in case severe tailing is observed on falling edge of
the
near end reflection peak of the test trace:
l
Use absorbent cotton ball with proper amount of
absolute ethyl alcohol to clean the tested fiber end and
optical output port end of OTDR.
l
Adopt smaller pulse width to retest the trace (note:
please select average test mode and set certain average
time here).
l
Under the pulse width currently set, use larger
attenuation to retest the trace (note: please select average
test mode and set certain average time here).
Ø
Distance between event points on test trace is too small.