Io
SENSOR
Section 4 RF/ID Tag Reader
SMT-200 Operations and Technical Manual 2.00/R1
51
Using the SMT-200 keypad, enter the required number. This number will be
displayed under the text
String No
. If you make an error you can use the left
arrow to erase the last digit. If the default number is the string number you
want to assign confirm by pressing
ENTER/START
.
The highest possible string number is 99999.
You can now read tags one after another. If a tag is read that is already
stored in the tag list the corresponding string number will be displayed. By
pressing
ENTER/START
you can change this string number.
4.3.2
Test the strings
To tests geophone strings enter
Do a test...
in the
Main menu
.
An overview is displayed of the
string configuration. Bypass this
screen by pressing
ENTER/START
.
To start a geophone test
connect the string to the
SMT-200. Aim the tag reader at
the tag and pull the trigger. If
the tag is known the
corresponding string number is
displayed and the geophone
test is started.
For security and quality control reasons, it is not possible to change the string number from this menu.
However, should a ‘new’ tag be read a string number must be allocated to it. A screen just like the
‘read tags’ screen (as described earlier) pops up and the default number can be accepted or another
string number allocated. Confirm the string number by pressing
ENTER/START
, the tag ID and string
number is stored in the tag list and the test is started.
Should a string be encountered without a tag a string number can be manually allocated via the
SMT-200 keypad.
Read tags
----------------
Tag
1000103306812
String No.
2
NEW
----------------
Read next Tag
Typ new Str. No.
1/99999
Main menu
----------------
1>Do a test...
2 Setup menu...
3 Temp. menu...
4 File menu...
5 Tag menu...
6 About...
0 Shut down
Enter test mode
String config.
----------------
[Sensor-SM4]
Temp [20.0 °C]
Series [1]
Parallel [1]
Not shunted
Next# [1 A]
Estimated space:
3276 Records
----------------
0 Back to Main
- Repeat Test
Start next Test
--Test results--
No. 1
Temp 20.0 °C
Res 382 ohm
Freq 9.91 Hz
Damp 0.259
Sens 28.7 V/m/s
Dist 0.04 %
Imp 2205 ohm
Test OK
0 Back to Main
- Repeat Test
Start next Test