STP-H803/H1303 Series Instruction Manual
9.3
Parameter Set Mode
Parameter Set Mode is used to set the different parameters of the STP pump and
STP control unit. The following items can be set in Parameter Set Mode:
(1) Remote mode
(2) Rated speed
(3) TMS function enable/disable
(4) Inhibit function enable/disable
(5) Emergency vent. valve function enable/disable
(6)
Remote mode configuration
O
pe
ra
tio
n
◇
The emergency vent. valve is not used with this series of STP pump.
<Parameter Set Mode> (See Figure 9.3)
1) Push the "SELECT" and "UP" switches at the same time to enter Parameter
Set Mode.
The available menu items (items (1) through (6) above) are displayed in order.
2) Press the "UP" switch to display the next parameter and the "DOWN" switch
to display the previous one.
3) Press the "ENTER" switch to set the parameter to the new setting.
4) Press the "SELECT" switch to display the next item.
5) Change the parameter by pressing the "UP" and "DOWN" switches, then set
it by pressing the "ENTER" switch.
6) When the save screen is displayed, select "YES" then press the "ENTER"
switch to store the new parameters. "IN PROGRESS" is displayed on the
LCD followed by "STORAGE OK", when the new parameters have been
stored successfully, or "STORAGE NOT OK", when the new parameters have
been stored unsuccessfully.
7) Press the "SELECT" switch to return to Operation Mode.
Or, do not press any switch for approx. 1 minute and the display will
automatically revert to Operation Mode.
8) Select "NO" and press the "SELECT" switch to return to Operation Mode
without saving the new parameters.
9-6