SPX100 SENSOR
DUST EMISSIONS CONTROLLER
Notice Technique
FI 72.0110.0316E
Page 22 sur 27
16) Alarm adjustment / alarm reading
- The alarm adjustment is allowed if the password is good.
- To access this, keep the “alarm” key pressed :
- Select the digit you want to adjust with the “right” and “left” keys and press the “up” and
“down” keys to increase or decrease the selected digit.
- If the alarm is activated, the display shows :
TO CANCEL THE ALARM, PRESS THE “VALID” KEY
NOTE : - If the alarm is resetted, the averaged measure is initialised with the
instantaneous measure. So, if the instantaneous measure is upper than the alarm threshold,
the alarm is activated again.
- The alarm is stored into memory even if the supply drops and it is keeping
maintained (even if the measure drops below the alarm threshold).
- For operation with a "non maintained" alarm, please consult us.
BE CAREFUL
: The measure full scale is within the calibration. The alarm threshold
must be lower than the full scale. The maximum alarm threshold is displayed during the
alarm threshold adjustment by a rolling text. If the adjusted value is upper than the maxi,
the alarm adjustment mode is maintained until a good value.