Manual revision 001
Appendix II: Command Summary
SBE 39
plus
-IM
87
FUNCTION
CATEGORY
COMMAND
DESCRIPTION
39plus-IM
Acquisition
Microcontroller
Commands
(
ii
= 39
plus
-IM ID)
Status
#iiGetCD
Get and display configuration data.
#iiGetSD
Get and display status data.
#iiGetCC
Get and display calibration coefficients.
#iiGetEC
Get and display event counter data.
#iiResetEC
Reset event counter.
#iiGetHD
Get and display hardware data.
#iiHelp
Display list of currently available
commands.
#iiDS
Get and display status and
configuration data.
#iiDC
Get and display calibration coefficients.
General Setup
#iiDateTime=
mmddyyyyhhmmss
Set real-time clock month, day, year,
hour, minute, second.
#iiOutputExecutedTag=
x
x=Y:
Display XML Executing and
Executed tags.
Default
.
x=N:
Do not.
Memory Setup
#iiInitLogging
Initialize logging to make entire
memory available for recording.
#iiSampleNumber=x
x=
sample number for last sample in
memory.
#iiSampleNumber=0
equivalent to
#iiInitLogging
.
Output Format
Setup
#iiOutputFormat=x
x=0
: output raw decimal data.
x=1
: output converted decimal data.
Default
.
x=2
: output converted XML decimal
data.
#iiFormat=x
x=0
or
1
: Date and time format
dd mmm yyyy, hh:mm:ss (date before
time; 3-letter month abbreviation) if
#iiOutputFormat=0
or
1
.
Default
.
x=2:
Date and time format
hh:mm:ss , mm-dd-yyyy (time before
date; 2-digit month number) if
#iiOutputFormat=0
or
1
.
#iiOutputTemp=x
x=Y:
Output temperature.
Default
.
x=N
: Do not.
#iiSetTempUnits=x
x=0
: Temperature °C, ITS-90.
Default
.
x=1
: Temperature °F, ITS-90.
#iiOutputPress=x
x=Y:
Output pressure.
Default
.
x=N
: Do not.
#iiSetPressUnits=x
x=0
: Pressure dbar.
Default
.
x=1
: Pressure psi.
#iiTxSampleNum=x
x=Y:
Output sample number with data
from
!iiData
,
Dataii
, or
!iiGetReply
.
Default
.
x=N
: Do not.
#iiSetCoastal=x
x=0
: Set units to °C and dbar, enable
temperature and pressure output, disable
sample number output.
x=1
: Set units to °C and psi, enable
temperature and pressure output, disable
sample number output.
#iiLegacy=x
x=0
: Allow all commands.
Default
.
x=1
: Set units to °C and dbar, enable
temperature and pressure output, disable
sample number. Do not allow user to
change those settings. Modify
#iiDS
response to match firmware < 4.0, for