Manual revision 001
Section 3: Preparing SBE 39
plus
-IM for Deployment
SBE 39
plus
-IM
20
Setup of IMM for use with SBE 39plus-IM
Sea-Bird recommends
one
of the following setups of the IMM for use with
the SBE 39
plus
-IM:
To make full use of IMM capabilities:
*Init
*Init
(Resets IMM to factory default state [must be sent twice])
SetConfigType=2
SetEnableAutoIMFlag=0
SetEnableBackSpace=1
SetEnableBinaryData=0
SetEnableEcho=1
SetEnableHostFlagConfirm=0
SetEnableHostFlagTerm=0
SetEnableHostFlagWakeup=0
SetEnableHostPromptConfirm=0
SetEnableHostServeOnPwrUp=1
SetEnablePrompt=1
SetEnableHostWakeupCR=0
SetEnableSignalDetector=0
SetTermFromHost=36
SetTermToHost=13
To have the IMM emulate the SIM:
*Init
*Init
(Resets IMM to factory default state [must be sent twice])
SetConfigType=1
SetEnableBinaryData=0
Verify that the IMM is set up as described by sending the
GetCD
command
before proceeding with the SBE 39
plus
-IM setup.
A script including one of the above setups should be included in the buoy
controller, to allow the IMM to be reset to the appropriate state if it
becomes corrupted.
Note:
See the IMM manual for complete
details on the use and setup of the
IMM.