Manual revision 001
Section 4: Deploying and Operating SBE 39
plus
-IM
SBE 39
plus
-IM
46
Acquisition Microcontroller Status
Command
#iiGetCD
Get and display configuration data, which
includes parameters related to 39
plus
-IM setup.
Most of these parameters can be user-
input/modified. List below includes, where
applicable, command used to modify
parameter:
•
Device type, Serial number
•
Pressure sensor installed?
•
Output data format [
#iiOutputFormat=
]
•
Output temperature with each sample
[
#iiOutputTemp=
]?
•
Temperature units [
#iiSetTempUnits=
].
•
Output pressure with each sample
[
#iiOutputPress=
]? Only appears if
39
plus
-IM includes pressure sensor.
•
Pressure units [
#iiSetPressUnits=
].
Only appears if 39
plus
-IM includes
pressure sensor.
•
Output sample number with
!iiData
,
Dataii
, or
!iiGetReply
output
[
#iiTxSampleNum=
]?
•
Interval between samples for autonomous
sampling [
#iiSampleInterval=
]
Example:
SBE 39
plus
-IM with ID=03 with a pressure sensor (user input in bold, command used to modify parameter in parenthesis).
#03GETCD
<ConfigurationData DeviceType = 'SBE39plus-IM' SerialNumber = '03907844>
<PressureInstalled>yes</PressureInstalled>
(inclusion of pressure sensor set at factory)
<SampleDataFormat>converted engineering</SampleDataFormat>
[
#iiOutputFormat=
]
<OutputTemperature>yes</OutputTemperature>
[
#iiOutputTemp=
]
<TemperatureUnits>Celsius</TemperatureUnits>
[
#iiSetTempUnits=
]
<OutputPressure>yes</OutputPressure>
[
#iiOutputPress=
]
<PressureUnits>Decibar</PressureUnits>
[
#iiSetPressUnits=
]
<TxSampleNumber>yes</TxSampleNumber>
[
#iiTxSampleNum=
]
<SampleInterval>15</SampleInterval>
[
#iiSampleInterval=
]
</ConfigurationData>
Notes:
•
#
iiGetCD
output does not include
calibration coefficients. To
display calibration coefficients,
use the
#iiGetCC
command.
•
Lines describing what
parameters to output only appear
if
#iiOutputFormat=1
or
2
. Raw
output (
#iiOutputFormat=0
) is
not affected by enabling /
disabling parameter outputs.