Manual revision 001
Appendix II: Command Summary
SBE 39
plus
-IM
86
Appendix II: Command Summary
FUNCTION
CATEGORY
COMMAND
DESCRIPTION
Remote SIM
Commands
-
PwrOn
Send wakeup tone to
all
IMs.
PwrOff
Send power off command to
all
IMs.
IMs enter quiescent (sleep) state. Main
power turned off, but data logging and
memory retention unaffected.
DS
Display SIM firmware version
and status.
Baud=x
x=
baud rate between SIM and
computer/controller (1200, 2400, 4800,
or 9600).
Default 9600.
DataNNMax=x
x=
timeout (0-32767 msec) that applies
to
!iiData
and
Dataii
only. If no reply
received within
DataNNMax=
, control
returned to computer and other
commands can be sent.
Default 1000 msec.
RelayMax=x
x=
timeout (0-3276 sec) that applies to
all other commands. If no reply received
within
RelayMax=
, control returned to
computer and other commands can be
sent.
Default 20 sec.
EchoOn
Echo characters received from computer
(default).
EchoOff
Do not echo characters received
from computer.
AutoPwrOn=x
x=Y
: Automatically send
PwrOn
to
IMs when power applied to SIM,
waking up all IMs on-line.
Default.
x=N
: Do not send
PwrOn
when power
applied to SIM.
Integrated IMM
Commands
(
ii
= 39
plus
-IM ID)
ID and Group
Number
ID?
Get 39
plus
-IM ID (0-99).
*ID=ii
(only 1 IM can be on line
or all will have same ID)
Set IM ID to
ii
(0–99). Must be sent
twice, because verification requested.
!iiSetGroupNumber=x
x=
39
plus
-IM group number (0-9).
Group 0 is pre-defined as group of all
instruments with integrated IMM.
Status
!iiGetCD
Display integrated IMM configuration
data.
!iiGetHD
Display integrated IMM hardware data.
!iiGetSD
Display integrated IMM status data.
Testing
!iiTestCableCoupler
Test integrity of integrated IMM. Line
must be captured before command is
sent.
Get Data
!iiSetGDataStr=x
x=
character string to send to 39
plus
-IM
acquisition microcontroller from
integrated IMM when
GData
is sent
from remote IMM/SIM.
GData
Make time synchronized measurements
across multiple instruments of different
types. Command initiated when
GData
is sent is set with
!iiSetGDataStr=
.
Individual instruments hold response in
buffer until user sends
!iiData
,
Dataii
,
or
!iiGetReply
.
!iiData
or
Dataii
or
!iiGetReply
Get data in buffer, obtained with
GData
, from IM with ID=ii.
Notes:
•
See
Command
Descriptions
in
Section 4:
Deploying and
Operating
SBE 39plus-IM
for
detailed
information and
examples.
•
If using the
39
plus
-IM with a
remote IMM: see
the IMM manual
for IMM command
descriptions.