Manual revision 001
Appendix II: Command Summary
SBE 39
plus
-IM
88
FUNCTION
CATEGORY
COMMAND
DESCRIPTION
Continued . . .
39plus-IM
Acquisition
Microcontroller
Commands
(
ii
= 39
plus
-IM ID)
Autonomous
Sampling
(Logging)
#iiSampleInterval=x
x=
interval (5 - 21600 sec) between
samples. When commanded to start
sampling, 39
plus
-IM takes sample,
stores data in FLASH memory, and
powers down at
x
sec intervals.
#iiStartNow
Start logging now.
#iiResumeLogging
or
#iiStartInterval
Start logging in
#iiSampleInterval
sec.
#iiStartDateTime=
mmddyyyyhhmmss
Delayed logging start: month, day, year,
hour, minute, second.
#iiStartLater
Start logging at delayed logging
start time.
#iiGetAvgRestart
Get and transmit
average
data, start
next averaging cycle, and (re)start
logging in (
#iiSampleInterval
/ 2) sec,
#iiGetAvg
Get and transmit
average
data and start
next averaging cycle. Do not reset
logging time base.
#iiSS
Get and transmit raw statistics – number
of samples in average, and maximum,
minimum, delta, and average for
temperature, pressure, and pressure
temperature. Continue averaging.
#iiGetLastRestart
Get and transmit
last
data, and (re)start
logging in (
#iiSampleInterval
/ 2) sec.
#iiGetLast
Get and transmit
last
data. Do not reset
logging time base.
#iiGetNew
Get (take) and transmit new sample.
Data is
not
stored in FLASH memory.
#iiDNx
Upload last
x
scans (1–250) from
memory
while sampling
autonomously
.
#iiStop
Stop logging or waiting to start logging.
Must send
#iiStop
before uploading
data with
#iiGetSamples:b,e
.
Polled Sampling
#iiTS
Take sample and transmit data.
#iiTSR
Take sample and transmit
raw
data.
#iiTSS
Take sample,
store in FLASH
memory
, transmit data, and turn power
off.
#iiTSSOn
Take sample,
store in FLASH
memory
, and transmit data.
#iiTSN:x
Take
x
samples (1–1000) and transmit
data.
#iiSL
Transmit data from last sample from
buffer.
#iiSLT
Transmit data from last sample from
buffer, then take new sample
Data Upload
#GetSamples:b,e
Upload data from scan
b
to
e
. Send
#iiStop
before sending.
Testing
#iiTT
Measure temperature for 100 samples,
output converted data.
#iiTP
Measure pressure for 100 samples,
output converted data.
#iiTTR
Measure temperature for 100 samples,
output raw data.
#iiTPR
Measure pressure for 100 samples,
output raw data.
Note:
Use Upload menu
in SeatermIM or
Upload button in
SeatermUSB –
SBE39plus-IM to
upload data that
will be processed
by Plot39 or SBE
Data Processing.
Manually entering a
data upload
command does not
produce data with
the required format
for processing by
those programs.