Manual revision 001
Section 4: Deploying and Operating SBE 39
plus
-IM
SBE 39
plus
-IM
63
#iiOutputFormat=1 (default): Converted Decimal Data
(if
#iiFormat=0
or
1
)
ii, nnnnn,tttt.tttt,ppppp.ppp, dd mmm yyyy, hh:mm:ss, sample, navg
or
(if
#iiFormat=2
)
ii, nnnnn,tttt.tttt,ppppp.ppp, hh:mm:ss, mm-dd- yyyy, sample, navg
where
•
ii = 39
plus
-IM ID (0 – 99); sent only in response to
!iiData
or
Dataii
.
•
nnnnn = 39
plus
-IM serial number (5 digits preceded by a space, omits
039
prefix); sent only in response to
!iiData
,
Dataii
,
!iiGetReply
,
#
iiGetAvg
, #
iiGetAvgRestart
, or polled sampling command (but not sent
for
#iiDNx
).
•
tttt.tttt = temperature; sent if temperature output enabled; °C or °F..
•
ppppp.ppp = pressure; sent if pressure sensor installed and output enabled;
dbar or psi. Number of digits to left of decimal place is dependent on
pressure sensor range.
•
dd mmm yyyy = day, month (3-letter abbreviation), year.
•
mm-dd- yyyy = month (2-digits), day, year.
•
hh:mm:ss = hour, minute, second.
•
sample = sample number (number of samples in FLASH memory at time
command to take a sample was sent); sent only if
#iiTxSampleNum=Y
and in response to
!iiData
,
Dataii
,
!iiGetReply
, #
iiGetAvg
,
#
iiGetAvgRestart
, or a polled sampling command (but not sent for
#iiDNx
).
•
navg = number of data samples contained in average; sent only in
response to
!iiData
,
Dataii
,
!iiGetReply
, #
iiGetAvg
, or
#
iiGetAvgRestart
.
Leading zeroes are suppressed, except for one zero to the left of the decimal
point. All data is separated with a comma.
Example:
Response to
!iiData
for 39
plus
-IM with
#iiOutputFormat=1
,
#iiFormat=0
, ID=03, pressure sensor installed,
#iiTxSampleNum=Y
:
03,09999, 8.5796, 531.316, 14 Jul 2016, 09:01:44, 1126, 250
(ID, serial number, temperature, pressure, date, time, sample number, number of samples in average)