System Installation and Wiring
For system installation and wiring details, refer to:
•
Mooring Cable and Wiring Requirements
in
Section 2: Description
of SBE 39plus-IM
•
Inductive Modem Module Manual
•
Appendix III: SIM Hookup and Configuration
.
Installing Optional Inductive Cable Coupler (ICC)
1.
Loosen the titanium hex head bolts connecting the two halves of each of
the ICC brackets. Pull the halves apart.
2.
Place the insulated mooring cable inside the brackets’ grooves.
3.
Reinstall each bracket half with the hex bolts.
4.
Verify that the two halves of the modem coupling toroid have come
together evenly, and that the mounting clamp is secure.
Note:
See
Application Note 85:
Handling of Ferrite Core on
Instruments with Inductive
Modem Telemetry
for more
detailed information on handling
and installation.
Wiring to
IMM/SIM
PCB
Mounting clamp, with opening
sized to specified cable
diameter – cable
clamped
by
this bracket.
Note: Installing clamps on
larger cable than specified
may cause damage to cable
and/or modem and prevent IM
communications.
Mounting guide /
inductive modem coupler
– contains modem
coupling toroid. Cable
goes through here but is
not clamped
, to avoid
putting through tension
on end cap (which could
pull off end cap).
Detail of guide and
core is similar to
shown above for the
39plus-IM guide and
core.