Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
When modifying this parameter with the “
set
” command, you must enter at
least the first two digits of one of these baud rates as the value parameter.
IMPORTANT! Make sure that each device on the network, including the
Network Master, is operating with the same network baud rate. Any
devices in the network running at a baud rate different from the Network
Master will be ignored.
Ideally, you would want to run the network at the fastest baud rate available.
However, certain restrictions, like cable quality or excessively long transmission
mediums, may require a reduction in the network baud rate.
O
PTICAL
D
ATA
A
VERAGING
E
NABLE
Command Line Parameter:
avg
Value Parameter:
on|off
This parameter enables or disables optical data averaging. The number of
ADC samples used for averaging is fixed depending on the
maxrate
parameter as follows:
Frame Rate (Hz)
Number of
Samples Averaged
0.125
60
0.25
30
0.5
15
1
7
2
3
4
1
8
1
10
1
12
1
0 (AUTO)
1
When modifying this parameter with the “
set
” command, you must enter either
“
on
” or “
off
” as the value parameter to enable or disable averaging.