Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
N
ETWORK
M
ODE
Command Line Parameter:
netmode
This parameter enables or disables network operation for the instrument.
Although disabling this parameter will force the instrument to run autonomously,
enabling it does not necessarily mean network operation will be invoked. The
instrument’s operating mode is determined during the initialization sequence.
See
Network Operation
in section
D - OPERATION
for more information.
When modifying th
is parameter with the “
set
” command, you must enter either
“
on
” or “
off
” as the value parameter to enable or disable network operation.
N
ETWORK
A
DDRESS
Command Line Parameter:
netadd
The network address uniquely identifies an instrument on a network. All
network transmissions use this parameter to identify the sender and receiver of
the message. It is not important what value is assigned to the network address,
as long as it is unique from other instruments in the network.
IMPORTANT! Make sure that each device on the network, including the
Network Master, has a unique network address. If two or more devices
have the same address, contentions may result and data could be lost.
When modifying this parameter with the “
set
” command, you must enter an
integer from 1 to 255 inclusive as the value parameter.
N
ETWORK
B
AUD
R
ATE
Command Line Parameter:
netbaud
The network baud rate defines the speed at which data is transferred on the
network interface. Baud rates are specified in units of
bits per second
(bps).
Only certain standard parameter values are accepted, as shown in the table
below:
Baud Rate (bps)
9600
14400
19200
28800
38400
57600
76800