Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
there is a space between the command and the “
-?
“ parameter. If there is
additional help available for the command, the text will be displayed. Otherwise,
a message indicating, “
No more help is available.
” will be displayed.
Some commands require additional command line parameters. Executing one
of these commands with missing or incorrect parameters will display a
“
Usage:
” message. This is helpful in determining what parameters are
acceptable for a particular command and how they should be formatted. For
example, if you executed the “
power
” command without any parameters, the
following message would be displayed:
Usage: power [operational power (on|off|?)]
This command requires one parameter, as indicated by the contents of the [ ]
brackets. If the command required more than one parameter, additional sets of
[ ] brackets with their parameter descriptions would be displayed. Parameters
must always be separated by a space. Within the [ ] brackets is a description
of the parameter followed by the list of acceptable parameter values, contained
in the ( ) brackets. The values listed here are always separated by the “
|
”
character. In this case, there are three accepted forms of this command;
“
power on
”, “
power off
”, and “
power ?
”.
Generally, the usage of the command console is self-explanatory. It should only
take you a few moments to get a working knowledge of the system. Although
the on-line help is fairly extensive, some commands need more detailed
explanations that would be too cumbersome to include in the instrument itself.
The following sections describe each command in more detail.
R
ESET
C
OMMAND
The “
reset
” command resets the console, redisplaying the command prompt
header described above. Any configuration parameters modified during the
console session that were not saved will revert back to their previous values.
This command requires no additional command line parameters.
ID
C
OMMAND
The “
id
” command displays the identification banner for the instrument, as
shown is the following example:
Sea-Bird Scientific OCR-504 Multispectral Radiometer
Copyright (C) 2011, Sea-Bird Scientific. All rights
reserved.
Firmware version: 5.1.0 - SatNet Type B
Instrument: SATDI4
S/N: 0001
The identification banner is also part of the start-up banner, which is displayed
during the initialization sequence described in section
D - OPERATION
. This
command requires no additional command line parameters.