Operation Manual
SYSTEM
OCR-504 Multispectral Radiometer
SECTION
E - CONFIGURATION
E
XIT AND
E
XIT
!
C
OMMANDS
The “
exit
” and “
exit!
” commands end the current command console
session. Once the console exits, normal operation will resume in most cases.
Otherwise, the instrument will reset itself before normal operation can begin.
The only difference between the two versions of this command is that the
“
exit!
” command forces a reset of the instrument, even if it isn’t necessary.
There are two conditions that will cause the instrument to reset itself. One or
both conditions must exist for this to occur. These conditions are:
1. The command console was invoked during the initialization sequence.
2. Configuration parameters have been modified and saved.
If you attempt to exit the console with modified configuration parameters that
have not been saved, the following dialog will be presented:
The configuration parameters have been modified.
Save changes [y/n]?
Choose “
y
” for yes or “
n
” for no to answer this question. If you choose yes, the
configuration parameters will be saved and the instrument will reset itself.
Otherwise, any modifications to the configuration parameters will be lost. See
the
Save Command
section above for more information on saving
configuration parameters.
OCR-504 Configuration Parameters
This section describes, in detail, the function of each configuration parameter
used by the OCR-504. The title of each section identifies the name of the
parameter, as displayed by the “
show
” command. Also clearly identified in each
section is the command line parameter keyword used in both the “
set
” and
“
show
” commands.
See the descriptions of the “
set
” and “
show
” commands described in the
Command Console
section above for more information.
S
ENSOR
T
YPE
Command Line Parameter:
sensortype
Value Parameter:
irradiance|radiance
The
sensortype
parameter defines whether the sensor is physically
configured as an irradiance or radiance sensor. This parameter is primarily
used to determine the instrument header in the telemetry frame (see the
Telemetry Format section). This setting can not be modified by the user, but
can be viewed with the
show
command.
T
ELEMETRY
F
RAME
T
YPE
Command Line Parameter:
frametype
Value Parameter:
binary|short|long