![SCI MM-380A Operation Manual Download Page 86](http://html1.mh-extra.com/html/sci/mm-380a/mm-380a_operation-manual_1217109086.webp)
MM-380A
12. Data Communication
12-4
Code
Item
09
Mean Force 1
10
Mean Force 2
11 Peak
Force
12 -
13 Force
Time
14 Mean
External
Voltage
15 Peak
External
Voltage
16 -
17 External
Time
18 -
19 -
Judgment Table
Code
Judgment
(Space)
No judgment (20H)
O
Overrange error
I
Pulsation error
2)
Waveform
Data string:
!01,00056:001,11/14 18:41:55,00, ,1.86kA,01, ,3.63kA,04, ,10.0CYC,
A
B C
11, ,0N,13, ,0ms[CR][LF]
0,0,2,4,[CR][LF],
D E F G
0000.0,-0.02kA, 0.0V, 0.00kW, 0.00N[CR][LF]
H I J K L
0000.5,-0.02kA, 0.0V, 0.00kW, 0.00N[CR][LF]
0001.0,-0.02kA, -0.1V, 0.00kW, 0.00N[CR][LF]
0001.5, 0.00kA, -0.1V, 0.00kW, 0.00N[CR][LF]
…
No output is produced if waveform display is OFF in the VIEW screen.
The data is as follows if the 4
th
item is OFF:
!01,00056:001,11/14 18:41:55,00, ,1.86kA,01, ,3.63kA,04, ,10.0CYC,
11, ,0N,13, ,0ms[CR][LF]
0,1,2,[CR][LF]
0000.0,-0.02kA, 0.0V, 0.00kW[CR][LF]
0000.5,-0.02kA, 0.0V, 0.00kW[CR][LF]
0001.0,-0.02kA, -0.1V, 0.00kW[CR][LF]
0001.5, 0.00kA, -0.1V, 0.00kW[CR][LF]
…
Item
Length
Description
A
Instrument No.
2 fixed digits from 00 to 31
Instrument-specific