MM-380A
8. Operation Screens
8-18
No.
Item
Description
(2)
Parameter
setting
TRIGGER
/MODE
(Continue
d from
previous
page)
(Continued
to next
page)
MODE: REAL TIME
Force constant start mode
Force is measured at intervals of half a
second.
In this trigger mode, the instrument
measures only force. To use this trigger
mode, select Constant Force (REAL TIME)
in the VIEW screen. Press the HOLD
button to put the instrument in hold mode
and stop the measurement.
<When TRIGGER: External (EXTERNAL) is
selected
*3
>
MODE: NORMAL
External
±
10 V voltage input normal mode
The instrument starts measurement each
time a force signal is input, making an
OK/NG judgment and showing the
measured value and waveform.
To pass multiple current pulses, refer to the
PULSE MODE settings in the SETUP (2)
screen.
The internal processing time varies depending on the force and
current flow time. As a result, the internal processing may not be
complete before next measurement.
*1:
Force when FORCE is selected, and external
±
10V voltage input when External
(EXTERNAL) is selected, for Measurement Selection (SELECT MEAS) in the
STATUS (1) screen.
*2:
Only when FORCE is selected for Measurement Selection (SELECT MEAS) in
the STATUS (1) screen.
*3:
Only when External (EXTERNAL) is selected for Measurement Selection
(SELECT MEAS) in the STATUS (1) screen.