8-5
No.
Item
Description
(1)
Measurement
(MEAS) 1 to 5
(Continued
from previous
page)
REAL TIME FORCE
The
MM-380A
constantly measures the force while the
hold is canceled. The
MM-380A
stops measurement
when put in hold mode. Measurement is made at
intervals of twice a second. Here, if you wish to select
REAL TIME FORCE for measurement, select FORCE
for TRIGGER and REAL TIME for MODE in the SETUP
(1) screen.
FORCE TIME
Shows the time from when the force signal exceeds the
force start level to when the signal falls below the force
end level.
EXTERNAL PEAK
Shows the mean external input voltage converted at the
preset conversion factor.
PEAK EXTERNAL VOLTAGE (P-EXT)
Shows the peak external input voltage converted at the
preset conversion factor.
REAL TIME EXT (R-EXT)
*1
The
MM-380A
constantly measures the external ±10
voltage while the hold is canceled. The
MM-380A
stops measurement when put in hold mode.
Measurement is made at intervals of twice a second.
Here, if you wish to select R-EXT for measurement,
select EXTERNAL for TRIGGER and CONSTANT for
MODE in the SETUP (1) screen.
EXTERNAL TIME
Shows the time from the start to the end of external
voltage input.
(2)
Waveforms
(WAVE)1 to 4
Select four items to display in the WAVEFORM screen
from the following:
CURR, VOLT, POWER, RESIST, FORCE, EXTERNAL
(3)
Display
ON/OFF
Select whether to show the waveforms by selecting
ON/OFF.
(4)
Function keys
Return
(BACK)
Pressing operation button E returns the display
to the previous screen (MEASUREMENT or
WAVEFORM screen).