![SCI MM-380A Operation Manual Download Page 37](http://html1.mh-extra.com/html/sci/mm-380a/mm-380a_operation-manual_1217109037.webp)
MM-380A
8. Operation Screens
8-11
h. ALL CYCLE (Force) Screen
No.
Item
Description
(1)
SCH.#
Shows the measurement condition number used (1 to
127). On the other hand, you can change conditions by
selecting this field with the encoder and turning the
encoder clockwise or counterclockwise. The details of the
conditions can be set using the SETUP (1) to (3) screens.
(2)
RMS per cycle
Shows the RMS FORCE every 10 ms. The area with
values shown in blue represents the calculation interval.
(3)
Measured
value
*1
Shows the measured values of five (5) items.
Measurement items can be selected in the VIEW screen.
Force
(CURR)
Pressing operation button A displays the
ALL CYCLE screen.
Recalculate
(RECALC)
Pressing operation button B recalculates
the measured force. Use this function to
redo the calculation of the mean over a
new calculation interval after changing the
start and end cycle (FORCE TIME, FORCE
TIMR END) settings in the SETUP (3)
screen.
(4)
Function keys
Scroll
(SCROLL)
Pressing operation button C and turning
the encoder clockwise or counterclockwise
scrolls the screen down or up.
*1:
Not shown if force is not measured.
(2)
(4)
(1)
(3)