8. Operation Screens
8-16
No. Item
Description
(2)
Parameter
setting
TRIGGER
/MODE
(Continue
d from
previous
page)
(Continued
to next
page)
<When TRIGGER: AUTO is selected
>
MODE: NORMAL
Auto normal mode
The first input among current, force (or
external
±
10V voltage input)
*1
and external
force trigger triggers measurement.
When current is a trigger
The instrument operates in the same
manner as in current or force in normal
mode.
When force (or external
±
10V voltage
input)
*1
is a trigger
The instrument operates in the same
manner as in force (or external
±
10 V
voltage input)
*1
normal mode.
When external force trigger is a trigger
The instrument operates in the same
manner as in force (or external
±
10 V
voltage input)
*1
normal mode. Note,
however, that in the case of external
force trigger, the instrument starts
measurement only upon input of current
or force (or external
±
10V voltage
input)
*1
.
When you have selected Auto for Trigger,
attach the force sensor. The instrument
performs a zero force adjustment when the
hold mode is turned off.
<When TRIGGER: FORCE is selected
*2
>
MODE: NORMAL
Force normal mode
The instrument performs measurement
each time a current signal is input, showing
the measured values and waveforms.
To pass multiple current pulses, refer to the
PULSE MODE settings in the SETUP (2)
screen.
The internal processing time varies depending on the force and
current flow time. As a result, the internal processing may not be
complete before next measurement.
*1:
Force when FORCE is selected, and external
±
10V voltage input when External
(EXTERNAL) is selected, for Measurement Selection (SELECT MEAS) in the
STATUS (1) screen.
*2:
Only when FORCE is selected for Measurement Selection (SELECT MEAS) in
the STATUS (1) screen.
*3:
Only when External (EXTERNAL) is selected for Measurement Selection
(SELECT MEAS) in the STATUS (1) screen.