Defibrillation
5
User guide
Semi-automated defibrillation
5.5
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
Step 3
Step 3 shock delivery
As soon as the energy for a shock is charged, the device prompts the user to deliver
the shock by pressing button 3.
Danger of electric shock!
• Do not, under any circumstances, touch the patient during shock delivery.
• Make sure that the patient does not touch any conducting objects.
9. Deliver the shock by pressing the button
.
After the shock delivery, step
4
follows.
Step 4
Cardiopulmonary resuscitation
10. Carry out cardiopulmonary resuscitation. Alternate between 30 chest compres-
sions and 2 breaths
1
for 2 minutes
2
. After 2 minutes, the device begins again with
.
11. Finish the therapy (see page
The following default energy values are programmed:
Shock
Adults
Children
1
150 joules
50 joules
2
200 joules
50 joules
3
200 joules
50 joules
1. A "continuous compressions" option is also available (i.e. no rescue breaths)
2. CPR cycle duration can vary depending on the "CPR cycle configuration" settings.
The CPR duration may vary according to country-specific standards (see page
Defibrillator ERC Protocol).