5
Defibrillation
5.5
Semi-automated defibrillation
Page 94
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
5.5.3
Defibrillation procedure
When the device is switched on, spoken and displayed instructions are issued
regarding the defibrillation. Closely follow the instructions.
Step 1
Switching on and preparing the device
Step 2
Analysis
1. Switch on the device by pressing the green button or the AED button directly.
2. Check the state of the patient.
3. Connect the electrode cable to the pads connector.
4. You are prompted to continue the resuscitation and to apply the electrodes.
5. Apply the defibrillation electrodes (see section
5.3.1 Applying the adult and pae-
, page
).
The message
CONNECT THE ELECTRODES
is switched off as soon as the de-
vice measures an acceptable electrode resistance. If it is not switched off, see
section
5.3.1 Applying the adult and paediatric electrodes page 85
.
Fig. 5.8
Switch unit on
6. The analysis starts automatically when the electrodes are detected.
7. You are prompted not to touch the patient any more.
8. The analysis key can be pressed any time during CPR to start a new analysis.
If the device detects ventricular fibrillation or ventricular tachycardia with a heart rate
exceeding 150 pulse/min,
follows; otherwise continue with
Cardiopulmonary resuscitation, page
Fig. 5.9
Analysis