Monitoring
4
User guide
ECG and heart rate monitoring
4.4
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
4.4.5
Starting ECG monitoring
1. Apply the electrodes as shown in
or
2. Connect the patient cable to the ECG signal input.
3. Define the ECG settings directly via the Touch screen curve or measurement
field.
4. Open the HR module (ECG menu) and check the settings.
Fig. 4.6
ECG cable
The screen
The sweep speed on the screen is fixed to 25 mm/s.
Curve List
ECG: Defi
ECG: I
ECG: II
ECG: III
ECG: aVr
ECG
HR source
Auto Scale
Amplitude
ECG filters
QRS sound level
QRS complexes are represented by green vertical dashes above the top ECG curve,
pacemaker pulses by red vertical dashes.
Curve List
ECG: II
ECG: Defi
Back to Default
Curve list with a 4-wire cable
Curve list with a 10-wire cable
Screenshot
Manual Def
Start
Menu
HR
---
Off
NIBP
Event
R-ECG
Adult
OFF