Page 131
Technical Data
12
User guide
System data
12.1
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12 Technical Data
12.1 System data
Data refer to standard testing conditions.
Technical data about the defibrillation (chapter
) applies only to the
DEFIGARD
®
Touch 7
.
Manufacturer
SCHILLER MEDICAL
Device type
DEFIGARD
®
Touch 7
&
PHYSIOGARD
®
Touch 7
Dimensions
160 x 250 x 70 mm (h x l x w) without bag
Weight
3.3 kg with battery and bag for
DEFIGARD
®
Touch 7
2.9 kg with battery and bag for
PHYSIOGARD
®
Touch 7
Protection case
IP 55
Power supply DC/DC
Input
Output
Ambulance bracket
10.8 till 17.6 VDC
15 VDC/4.0 A
Power supply AC/DC
(desktop
charger and ambulance bracket)
Input
Output
Type XP Power Model: AHM85PS15
Medical grade switching power supply, protection class I.
100 - 240 VAC, max. 1.0 A, 50-60 Hz
15 VDC, max. 5.67 A
Power battery
Battery type
Autonomy
Charging time
Safety primary cell
Battery type
Lithium/ion 11.1 V, 4.65 Ah, 51.6 Wh
100 shocks with maximum energy or >6 hour monitoring
90 %: 2 hours after full discharge and device switched off
Ensures continued monitoring for approx. 30 seconds when replacing the power
battery
Lithium/MnO2, 6 V, 1.4 Ah
Environmental conditions
For operation
• 0 ºC ... 40 ºC relative humidity at 15 - 95 % (non condensing)
Atmospheric pressure 700...1060 hPa
• If higher or lower temperatures prevail during use, a limited operation time of up
to
1 hour
is possible, if device has been stored previously at room temperature.
See next page “environmental conditions for transient operation”