4
Monitoring
4.9
Temperature monitoring
Page 62
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
4.9 Temperature monitoring
4.9.1
Start temperature monitoring
4.9.2
Temperature menu settings
Access the temperature menu via the TEMP measurement field as described on
.
The default settings are printed
bold
.
4.9.3
Temperature alarms
• Depending on the sensor type, the sensor can be applied to the ear, the skin or per
rectum.
• To achieve a reliable measured value, independent of the measuring site, the
measurement duration must be at least 2 minutes.
• The temperature measurement method is “Direct mode”.
1. Connect the sensor to the temperature input.
2. Select the TEMP measurement
field to open the Temp menu.
3. Select OFF/ON button to start the measurement.
Menu item
Parameter
Description
Start measurement
On or
OFF
Calibrate
Calibration of the sensor
Alarm
Cause
Remedy
Check sensor
• Temperature sensor not connect-
ed to the device
Connect sensor.
TEMP: Out of range
• Temperature is out of set alarm lim-
its.
Check patient
Check narrow/wide alarm limit and adjust it if neces-
sary.