12
Technical Data
12.6 Device Configuration
Page 150
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
12.6.5
AED
12.6.6
ECG
12.6.7
IBP
Parameter
Values
Description
Display curve in
AED
• False
• True*
If enabled, ECG lead Defi is displayed in AED mode
AED voice level
• High
• Medium*
• Low
Metronome default behaviour in adult mode
This behaviour can be changed during intervention on the device
Metronome
behaviour adult
• 30:2*
• 15:2
• Continuous Off
Metronome default behaviour in adult mode
This behaviour can be changed during intervention on the device
Metronome
behaviour child
• 30:2
• 15:2*
• Continuous Off
Metronome default behaviour in child mode
This behaviour can be changed during intervention on the device
Metronome rate
•
100cpm*
• 110cpm
• 120cpm
Sets the frequency of the metronome
Metronome audio
level
• High
• Medium*
• Low
Sets the audio level of the metronome
CPR depth unit
• cm*
• inch
Sets the unit in which the chest compression depth value will be
displayed
CPR cycle duration
• 1min
• 2min*
• 3min
Sets the duration of the CPR cycle
Parameter
Values
Description
ECG curve
amplitude
• 0.25mv
• 0.5mv
• 1mv*
• 2mv
• auto
Sets the default ECG curve amplitude, If Auto is selected, the
displayed amplitude will be automatically adapted depending on the
signal amplitude
ECG 16,7 Hz filter
•
False*
• True
Not used at the moment, has no impact on the unit
QRS sound level
• Off
•
Low*
• Medium
• High
Sets the audio level of the QRS sound issued by the ECG
Parameter
Values
Description
IBP curve amplitude
•
30
, 60, 100, 300mmHg
Sets the default IBP curve amplitude