6
Pacemaker
6.4
Start-up of the Pacemaker
Page 102
A
rt.
no.
: 0-48-
0227
Rev.:
g
DEFIGARD/PHYSIOGARD Touch 7
6.4 Start-up of the Pacemaker
Shock hazard!
Pacing is started immediately when the pacemaker is switched ON and the
current is set.
In order to be able to operate the pacemaker, the following conditions must be met:
• the pacemaker (optional) needs to be activated.
• Pads must be connected to the device.
• When the pacemaker is switched on, the current value is set to 10 mA.
• The device can be switched from pacing to defibrillation mode at any time. The
pacemaker is stopped by confirming the switchover.
• The device can be switched from pacing to monitoring mode at any time. In this
case, the pacemaker screen is displayed as small measurement field at the top
right.
• If the pacemaker function is
OFF
and closed by pressing the
Close key
, the
frequency and current settings are reset.
• Pressing the monitoring
or
Close key
will minimize the pacemaker
window (pacing is still running).
• When pacemaker is started only "Monitoring advanced" and "Critical care" views
are available.
OFF