Page 59
A
rt.
no.
: 2.
511
076 Rev.:
g
Maintenance
6
User Guide
Error Messages
6.5
BR-102 plus / BR-102 plus PWA
182
Signal
disturbed
• Too much interference in Ko-
rotkoff signal.
Perform measurement in quiet environment;
avoid moving the arm during measurement.
190
No signal
• Pressure reached 50 mmHg
and no pulse detected (OSC).
Check cuff.
191
Weak signal
• Less than 8 pulse beats de-
tected (OSC).
Check cuff.
192
Signal
disturbed
• More than 200 pulse beats
detected (OSC).
Avoid moving the arm during measurement.
198
DIA not
detected
• Minimum pressure reached
without diastolic value detect-
ed (OSC).
Avoid moving the arm during measurement.
199
SYS not
detected
• Minimum pressure reached
without systolic value detect-
ed (OSC).
Avoid moving the arm during measurement.
200
No pulse
• No pulse detected (OSC).
Repeat measurement; check cuff.
201
Pulse > max.
• Too many pulse beats detect-
ed (OSC).
Perform measurement in quiet environment;
avoid moving the arm during measurement.
202
Pulse < min.
• Too few pulse beats detected
(OSC).
Repeat measurement; check cuff.
210
No pulse
• No pulse detected (KOR).
Connect microphone; Check position of
microphone. Ensure that the microphone is
correctly orientated (metallic (yellow) side
towards arm).
211
Pulse > max.
• Too many pulse beats detect-
ed (KOR).
Perform measurement in quiet environment;
avoid moving the arm during measurement.
212
Pulse < min.
• Too few pulse beats detected
(KOR).
Place microphone over brachial artery;
ensure that the microphone is correctly
orientated (metallic (yellow) side towards
arm).
238
Min. pressure
• Minimum pressure reached
without measurement result.
Repeat measurement.
239
Max. pressure • Maximum pressure reached
without measurement result.
Repeat measurement.
255
Internal error
• Device is defective.
Contact technical service.
322
No
measurements
available
• No recording stored.
• No recording has been made.
• Incorrect initialisation of a re-
cording.
• Error in the recording storage.
Take/ retake a recording.
323
No patient data
available
• No patient data has been re-
corded.
Enter patient data.
324
SD- card
missing
• SD card not inserted. For
PWA recording an SD card is
necessary.
Insert an SD card.