Maintenance
6
User Guide
Measurement Check
6.4
BR-102 plus / BR-102 plus PWA
6.4.4
Overpressure Relieve Valve
The overpressure relieve valve can be checked for correct release for
both the adult settings and paediatric setting.
Adult
1. Set / check that Adult is set in the record setup. This can be done
for single measurement or new record:
2. Setup and enter the calibration screen as detailed on the previous
page.
3. Pump to 300 mmHg. Slowly continue to pump and check that the
overpressure valve activates (a pressure hiss is heard and the cuff
rapidly deflates) at
320 mmHg ± 10 mmHg
.
Paediatric
1. Set the record setup to Child. This can be done for single meas-
urement or new record:
2. Setup and enter the calibration screen as detailed on the previous
page.
3. Pump to 200 mmHg. Slowly continue to pump and check that the
overpressure valve activates (a pressure hiss is heard and the cuff
rapidly deflates) at
225 mmHg ± 5 mmHg
.
NEW RECORD
LAST RECORD
SINGLE MEAS.
SYSTEM
NEXT
OK
05:16PM
RECORD SETUP
PROGRAM
PATIENT ID
START
BACK
NEXT
OK
NEW RECORD
NEXT
OK
RECORD SETUP
PatGrp
ADULT
RecDur
24 hr
DefSpd
5 mmHg/s
DispResult
ON
BACK
NEW RECORD
LAST RECORD
SINGLE MEAS.
SYSTEM
NEXT
OK
05:16PM
RECORD SETUP
PROGRAM
PATIENT ID
START
BACK
NEXT
OK
NEW RECORD
NEXT
OK
RECORD SETUP
PatGrp
CHILD
RecDur
24 hr
DefSpd
5 mmHg/s
DispResult
ON
BACK
Summary of Contents for BR-102 PLUS
Page 6: ...Page 6 Art no 2 511076 Rev g BR 102 plus BR 102 plus PWA ...
Page 72: ......