Page 39
A
rt.
no.
: 2.
511
076 Rev.:
g
BP recording
3
User Guide
Long Term Recording
3.4
BR-102 plus / BR-102 plus PWA
3.4.5
Stopping the Recording
The recording will stop automatically after 24 or 48 hours and no user
intervention is required. The recorder can however, be stopped
manually if required. Do this as follows:
Press and hold left green
Function key
for 4 seconds and confirm
with
YES - blue right Function key
.
If no confirmation is received within 30 seconds, the recording
continues.
3.4.6
Displaying a Recording on the BR-102 plus
1. Select
Last Record
from the main menu.
2. Select
Measurements
to display all measurements.
3. The measurements are displayed giving:
– Date and time.
– Measurement method:
KOR = Auscultatoric (Korotkoff/Riva-Rocci)
OSC = Oscillometric
– Systolic Pressure in mmHg.
– Diastolic Pressure in mmHg.
– Mean Arterial Pressure (MAP) in mmHg.
– Pulse rate (PULS) in beats / minute (bpm).
4.
Next
continues to the following measurement and
Back
returns to
the previous measurement.
5. To return to the main menu press
Back
for approximately 4
seconds. The recorder also returns to the menu automatically
when no key is pressed for approximately one minute.
NO
YES
STOP RECORDING
Do you want
to stop the
BP recording?
OK
STOP RECORDING
BP recording
Stopped
NEW RECORD
LAST RECORD
SINGLE MEAS.
SYSTEM
NEXT
OK
05:16PM
NEXT
OK
LAST RECORD
NEW RECORD
LAST RECORD
SINGLE MEAS.
SYSTEM
NEXT
OK
05:16PM
RECORD DATA
PATIENT DATA
MEASUREMENTS
BACK
MEASUREMENTS
12.05.2013
10:19
KOR
SYS
128
99 mmHg
mmHg
DIA
86
71
bpm
MAP
BACK
NEXT
PULSE
Summary of Contents for BR-102 PLUS
Page 6: ...Page 6 Art no 2 511076 Rev g BR 102 plus BR 102 plus PWA ...
Page 72: ......