5 Cleaning
5.1 Important Information
5.1.1
Non-admissible detergents
Never use products containing the following:
• Ethyl alcohol
• Acetone
• Hexane
• Abrasive cleaning powder
• Plastic-dissolving products
• Some patients have intolerances (e.g. allergies) to disinfectants or
their components. If you have such a patient or you are not sure, re-
move possible residues with careful washing.
• Do not autoclave the unit or any accessories.
• Do not immerse the device in liquid. If liquid does penetrate the unit,
switch it off immediately and send it to SCHILLER for testing.
•
Never
use a wet or dripping cloth and
never
spray the equipment
with detergent. Otherwise, the detergent may seep under the edges
of the keyboard or penetrate the device and destroy the electronics.
• Use of cleaning solutions which have a high acid content or are oth-
erwise inappropriate can cause damage to the equipment, including
cracking and deterioration of the plastic case.
• Always follow the mixing/diluting instructions provided by the man-
ufacturer of the cleaning solution.
• The pressure tube and microphone cable must not be exposed to
excessive mechanical stress. Whenever disconnecting, hold the
plug/connector and not the cable.
• When cleaning, ensure that all labels and safety statements, wheth-
er etched, printed or stuck to the unit, remain in place and remain
readable.