1.10 Operation with other Devices
Only use accessories and other parts recommended or supplied
by SCHILLER AG. The use of other than recommended or
supplied parts may result in injury, inaccurate information and/or
damage to the device.
Accessory equipment must be certified according to the respective
IEC standards (e.g. IEC/EN 60950 for data processing equipment
and IEC/EN 60601-1 for medical equipment). Furthermore, all
configurations shall comply with the valid version of the system
standard IEC/EN 60601-1. Everyone who connects additional
equipment to the signal input part or signal output part configures
a medical system, and is therefore responsible that the system
complies with the requirements of the valid version of the system
standard IEC/EN 60601-1. If in doubt, consult the technical service
department or your local representative.
The BR-102 plus / BR-102 plus PWA is safe during defibrillation.
However, as a safety precaution remove the cable assembly
between the recorder and the PC and when possible, remove the
BR-102 plus from the patient before defibrillation.
The BR-102 plus / BR-102 plus PWA complies with EMC
regulations for medical products which afford protection against
emissions and immunity. However, the possibility exists that high
frequency disturbance from other devices can affect the recorder's
operation.