Introduction
2
User Guide
Menu Structure
2.6
BR-102 plus / BR-102 plus PWA
2.6.1
Menu Overview
Main Menu
Sub
Menu 1
Value/Info
Sub Menu
2
Value/Info
Sub
Menu 3
Value/Info
NEW
RECORD
RECORD
SETUP
Defines the
recording
parameters
Patient
Group
Adult / Child - Maximum initial pressure for adult is
300mmHg and for a child 210mmHg.
Record
Duration
24/48 Hours
Deflation
speed
2, 3, 4, 5, 6, 7, 8, 9 mmHg/s or Auto (3mmHg / heart
beat).
Display
Result
ON / OFF. Displays the cuff pressure during a measure-
ment and displays the result on the screen after a meas-
urement has been taken. When Off is selected measure-
ments are not displayed during, nor after a
measurement.
Note: When Off is selected, the first two measurements
taken are always displayed to enable checking. The rec-
ommended setting is Off to prevent patient concern dur-
ing the recording.
PROGRAM
Defines the
program
parameters
Select
program
A, B, C, D
Select one of four individual recording programs A, B, C,
D. The programs are defined in
Program Setup
below.
PRO-
GRAM
SETUP
A, B, C, D - Four
individual BP
measurement pro-
grams (A, B, C
and D) can be de-
fined. Each indi-
vidual program
has up to 8 cycles
with separate start
time BP measure-
ment interval for
each cycle.
Change / Next
1.Define the start time by press-
ing
change.
The time counts
up in 10 minute intervals.
2.When the desired start time is
shown press
next
. The
measurement interval is
highlighted.
3.Press
change
to select a BP
measurement interval of: 5, 10,
15, 20, 30, 60, or 120 minutes.
4.Press
next
to define the next
start time.
5.Continue until all desired start
times (with measurement
interval) are defined for the
selected program.
When less than 8 cycles are required
pressing next twice returns to the
previous menu.
PATIENT ID
Record
Voice
Start
Start Voice Recording.
Stop
Stop Voice Recording.