Maintenance
6
User Guide
Calibration
6.3
BR-102 plus / BR-102 plus PWA
6.3 Calibration
6.4 Measurement Check
A service option is incorporated in the BR-102 plus / BR-102 plus
PWA to check the measurement accuracy and the correct functioning
of the overpressure valve. This option can be performed at any time
to check the unit‘s integrity.
6.4.1
Equipment Required
6.4.2
Setup
The unit must be returned to a SCHILLER approved centre for
calibration at the interval defined by local regulations or at least every
two years. A reminder message is displayed 30 days before
calibration is due and counts down the subsequent days every time
the unit is switched on.
The message is displayed for approximately 60 seconds before the
main BR-102 plus / BR-102 plus PWA menu is displayed. Select OK
to display the main menu immediately for normal use.
Time limit
22 day(s)
Call SCHILLER
service!
ok
DEVICE INSPECTION
• Calibrated Manometer (purchased locally)
• BR-102 plus / BR-102 plus PWA connector and hose assembly.