Service Mode and Demo Mode
ARGUS VCM
Factory default settings are described in Table 8.
Table 8. Factory Default Settings for ARGUS VCM
Parameter
Ranges
Defaults
Systolic (mmHg) Neonatal
Low: 25 to 115
High: 30 to 120
50, 100
Alarm Limits
Pediatric
Low: 30 to 155
High: 35 to 160
75, 145
Adult
Low: 30 to 255
High: 35 to 260
75, 220
Diastolic
(mmHg)
Neonatal
Low: 10 to 100
High: 15 to 105
30, 70
Alarm Limits
Pediatric
Low: 15 to 125
High: 20 to 130
35, 100
Adult
Low: 20 to 250
High: 25 to 255
35, 110
MAP (mmHg)
Neonatal
Low: 10 to 105
High: 15 to 110
35, 80
Alarm Limits
Pediatric
Low: 15 to 135
High: 20 to 140
50, 110
Adult
Low: 20 to 250
High: 25 to 255
50, 120
PR (bpm)
Neonatal
Low: 25 to 295
High: 30 to 300
100, 200
Alarm Limits
Pediatric
Low: 25 to 295
High: 30 to 300
50, 150
Adult
Low: 25 to 295
High: 30 to 300
50, 120
SpO
2
%
Neonatal
Low: 50 to 98
High: 52 to 100
85, 98
Alarm Limits
Pediatric
Low: 50 to 98
High: 52 to 100
90, 100
Adult
Low: 50 to 98
High: 52 to 100
90, 100
Patient Type
Adult, Pediatric, Neonatal
Adult
NIBP Units
mmHg, kPa
mmHg
NIBP initial
cuff inflation
Neonatal
50, 60, 70, 80, 90, 100, 110, 120, 132 mmHg
(6.6, 7.9, 9.3, 10.6, 11.9, 13.3, 14.6, 15.9, 17.5 kPa)
90 mmHg
(11.9 kPa)
Pediatric
80, 90, 100, 110, 120, 130, 140, 150, 160, 170 mmHg
(10.6, 11.9, 13.3, 14.6, 15.9, 17.2, 18.6, 19.9, 21.2, 22.6
kPa)
120 mmHg
(15.9 kPa)
Adult
100, 120, 140, 160, 180, 200, 220, 240, 270 mmHg
(13.3, 15.9, 18.6, 21.2, 23.9, 26.6, 29.2, 31.9, 35.9 kPa)
160 mmHg
(21.2 kPa)
NIBP Auto Interval
Off, STAT, 1, 2, 3, 4, 5, 10, 15, 30,
45, 60, 90, 120, 240
15
Temperature Units/Modes
°C, °F, °C M, °F M
°F
Pulse Tone Volume
0 to 8
4
Alarm Volume
1 to 8
4
Night Panel
On, Off
Off
Print Control
Manual, Stream
Manual
28
ARGUS VCM
Service Manual