Disassembly Guide
ARGUS VCM
Prior to Disassembly
1. Turn
ARGUS VCM
off by pressing the Power On/Off switch.
2. Disconnect the monitor from the AC power source
Fuse Replacement
1. After
Step B2
, remove 2 AC main fuses (F1, F2: 250V/2A) out of the socket if required.
2. Remove 1 battery fuse (F3: 250V/6.3A) out of the socket if required.
3. Replace (a) new fuse(s).
4. Reassembly the monitor.
Battery Replacement
This section describes the steps to remove the battery from
ARGUS VCM
for replacement
without disassembling the main case of the monitor
B1. Battery disassembly
5. Remove 4 small flat-head screws (S8094-0) fastening Battery cover.
6. Remove Battery cover.
7. Remove Battery carefully from the monitor.
8. Disconnect the wire between 2 Battery terminals.
Figure 16. Battery Disassembly
ARGUS VCM
Service Manual 45