Performance Verification
ARGUS VCM
Temperature Operation
If Temperature option installed, the following test procedures will verify temperature
performance.
1. Connect the monitor to an AC power source.
2. Press Power button to turn the monitor on.
3. Connect temperature probe to the temperature connector of the monitor.
4. Set Temperature tester to 96.4
° F.
5. Press the mode button (setting mode) to select ‘F M’ (monitored mode).
6. Insert the temperature probe into ‘…’ of Temperature tester when ‘Ready’ indicator of
the test is on.
7. After 3 minutes, verify that the monitor displays 96.1
° F ~
96.7
° F.
Measurement accuracy:
±0.2° F
Temperature tester accuracy:
±0.1° F
8. Return the probe into the probe holder.
9. Verify that Temperature display indicates ‘- -‘.
20
ARGUS VCM
Service Manual