Service Mode and Demo Mode
ARGUS VCM
Night Panel
Night panel is used to adjust the light intensity of the display.
With ‘Night Panel’ selected,
- select ‘YES’ using only Up (+) selection button to degrade the light intensity between
21:00 to 06:00.
- select ‘No’ using only Down (-) selection button not to use this option.
Figure 10. Night Panel
NIBP Calibration
‘NIBP Calibration’ is used to calibrate NIBP pneumatic pump.
With ‘NIBP Calibration’ selected,
1. Connect the NIBP adult hose to the monitor.
2. Kink the distal end (farthest from the monitor).
3. Press Up (+) selection button to start calibration.
4. The monitor displays ‘ING’ during the calibration.
5. Wait for about 20 seconds to complete.
6. Verify that Time display indicates one of digits from 040000 to 070000 after ‘End’
message appears. (If fails, refer to
Troubleshooting
section.)
7. Turn off the monitor.
8. After a few seconds, power on the monitor.
Figure 11. NIBP Calibration
ARGUS VCM
Service Manual 29