Specification
ARGUS VCM
SpO
2
/Pulse Rate
%Saturation
Range
1% to 100%
Low Perfusion
0.03% to 20%
Accuracy
Without Motion-Adults 70% to 100% ±2 digits
1% to 69% unspecified
Without Motion-Neonate
1
70% to 100% ±3 digits
1% to 69% unspecified
Low Perfusion
2
70% to 100% ±2 digits
1% to 69% unspecified
Pulse Rate
Range
20 BPM to 300 BPM
Accuracy Without
Motion
2
20 BPM to 300 BPM ±3 digits
Low Perfusion
2
20 BPM to 300 BPM ±3 digits
Standards EN865:1997
1
Neonate specifications are shown for neonate sensors with ARGUS VCM. Saturation accuracy will vary
by sensor type recommended by the manufacturer..
2
Specification applies to monitor performance and was validated with Biotek and Nellcor simulators
Temperature
Probe Type
Thermistor probe
Range
26° C to 43° C (80° F to 110°F)
Display Accuracy
±0.1° C (±0.2° F)
Measurement units
° C, ° F
Measurement modes
Predictive, Monitored
Predictive Mode
One-time measurement in a single temperature reading which
is displayed at the end of the brief measurement period
Monitored Mode
Continuous measurement over an indefinite period.
Standards
ASTM E1112-00, EN12470-3
Printer
Type Thermal
Resolution 8
(dots/mm)
Printing speed
45 (mm/s)
Paper width
57 (mm)
ARGUS VCM
Service Manual 9