Troubleshooting
ARGUS VCM
SpO
2
Performance
The followings are symptoms of problems and recommended actions relating to SpO
2
operational performance. If the action requires replacement of a PCB or module, refer to
Disassembly Guide
.
1. No displays (blank) on SpO
2
display.
1-1. Replace SpO
2
module.
1-2. If problem persists, replace Main BD.
2. Dashes ‘- -‘ are displaying on SpO
2
display, but no data measured.
2-1. Verify that red LED of SpO
2
sensor is on. If not, replace the SpO
2
sensor.
2-2. Ensure that the LED intensity increases when opening the SpO
2
sensor.
2-3. If problem persists, replace SpO
2
module.
3. Error code, E49 displays.
3-1. Replace SpO
2
module.
Temperature Performance
The followings are symptoms and recommended actions to address problems with
temperature operational performance. If the action requires replacement of a PCB or
module, refer to
Disassembly Guide
.
1. No displays (blank) on Temperature display.
1-1. Check the connection of 5-pin wire between Temperature PCB and Main BD.
1-2. Replace Temperature PCB if problem persists.
1-3. Replace Temperature module if problem still persists.
1-4. If problem persists after Temperature module replaced, replace Main BD.
2. Dashes ‘- -‘ are displaying on Temperature display, but no data measured.
2-1. Use another temperature probe.
2-2. Check the connection of 5-pin wire between Temperature PCB and Main BD.
2-3. Replace Temperature PCB if problem persists.
2-4. Replace Temperature module if problem still persists.
3. Error codes, E63 and/or E64 often display(s).
3-1. Use another temperature probe.
3-2. Replace Temperature module if problem persists.
4. Error code, E65 often displays.
4-1. Replace Temperature PCB.
4-2. Replace Temperature module if problem persists.
5. Error code, E69 displays.
5-1. Replace Temperature module.
ARGUS VCM
Service Manual 39