69
MPS4264
Section 5: Software
Oven Variables (LIST O)
The oven settings control the thermal chamber when performing a multi-temperature calibration or validation of the
MPS4000. Reference Scanivalve Document: MPS Field Val if performing a single temperature validation with an SPC4000.
List Oven Settings (LIST O)
DESCRIPTION
Returns all of the oven settings
SYNTAX
LIST C
RETURNS
All of the oven settings followed by the prompt.
EXAMPLE
LIST O
SET IPOVEN 0.0.0.0 0
SET STARTOVEN 0
SET STOPOVEN 0
SET TEMPOVEN 0
>
Set the Oven IP Address, Port and Calibrator Number (IPOven)
DESCRIPTION
Sets the IP address and port for the oven When 0.0.0.0 0 0 is set no oven is used. This is only
applicable when using performing multi-temperature calibrations or validations.
SYNTAX
SET IPOVEN <ipadd> <port>
Where:
<ipadd> - Calibrator IP address
<port> - Calibrator port
RETURNS
Prompt when ready to accept a new command
EXAMPLE
SET IPOVEN 10.0.0.122 23 1
>
DEFAULT
0.0.0.0 0
RANGE
These values are not range checked.
Set the Oven Start Command (STARTOVEN)
DESCRIPTION
Sets the command to send the oven to start
SYNTAX
SET STARTOVEN [<start oven command>]
NOTE: If no command is entered, a CR is sent to the oven
RETURNS
Prompt when ready to accept a new command
EXAMPLE
SET STARTOVEN RUNM
>
DEFAULT
0
RANGE
These values are not range checked.
Summary of Contents for MPS4264
Page 1: ...MPS4264 Miniature Pressure Scanner Hardware and Software Manual Software Version 2 07...
Page 2: ......
Page 3: ......
Page 6: ...INTENTIONALLY LEFT BLANK...
Page 12: ...6 MPS4264 Section 1 Specifications INTENTIONALLY LEFT BLANK...
Page 18: ...12 MPS4264 Section 2 Introduction INTENTIONALLY LEFT BLANK...
Page 32: ...26 MPS4264 Section 4 Hardware INTENTIONALLY LEFT BLANK...
Page 38: ...32 MPS4264 Section 5 Software INTENTIONALLY LEFT BLANK...
Page 81: ...75 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 86: ...80 MPS4264 Section 6 Maintenance INTENTIONALLY LEFT BLANK...
Page 87: ...81 MPS4264 Appendix INTENTIONALLY LEFT BLANK...